Some compounds are very sensitive to water or oxygen or even nitrogen in the air.
In order to ensure that such compounds are not interfered by water and oxygen, special instruments and anhydrous and oxygen-free operation techniques must be used in the process of synthesis, separation, purification or analysis and identification
1.
For systems with less stringent requirements, a method of directly passing inert gas into the reaction system to replace the air can be used.
Second, the glove box
Sampling, weighing, grinding, etc.
There are simple glove boxes, such as those with plexiglass shells, and complex ones, such as buffer chambers, such as stainless steel shells (except for glass windows)
The glove box with a stainless steel shell is more expensive and consists of long neoprene gloves, an exhaust port, an air intake port and a well-sealed glass window
Advantages of the glove box: anhydrous and oxygen-free effect is good
Disadvantages of the glove box: inconvenient operation and unfavorable cable; all utensils and chemicals must be put in in advance, once forgotten, they need to restart, delaying the process or seriously affecting the results
3.
The Shrank line is also called an anhydrous and oxygen-free operating line, which is a set of inert gas purification and operating systems
Figure 23-17 is a schematic diagram of the Shrank line structure
Figure 23-17 Schematic diagram of the Shrank Line structure
Since commercially available inert gases generally contain oxygen and 10-50 ppm moisture, a purification treatment system is required
For the usual anhydrous and oxygen-free operation, it can be simplified according to actual needs, and generally only double-row pipes are required
Advantages: anhydrous and oxygen-free effect is good, and it is more convenient to use than a glove box
Disadvantages: complicated preparatory work
Suggestion: The double-row tube is relatively simple, and its application in general chemical laboratories can be widely promoted
.