    Solution to the deviation of the temperature and setting in the salt spray corrosion resistance test chamber

    • Last Update: 2022-09-30
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    The salt spray test chamber is used to detect the corrosion resistance of the metal or certain plating material surfaces and to evaluate the corrosion resistance
    of the metal or plating surface based on the results of the salt spray test chamber test.

    In the salt spray test, the corrosion resistance
    of the test material is studied by observing the corrosion of the test article.

    At present, there are three kinds of brine used in salt spray tests: neutral salt solution, acetate solution, and copper-accelerated acetic acid solution
    .

    The salt spray test chamber is used for corrosion resistance testing and film thickness testing of some metals, alloys, metal coatings, oxide coatings, organic coatings, color paints, varnishes, etc
    .

    When operating the salt spray test chamber, it can be operated by the temperature controller on the salt spray test chamber to adjust the temperature in the box and the temperature of the pressure barrel to meet the test requirements
    .

    In addition, the salt spray test chamber can also be used for corrosive gas testing, this type of salt spray test chamber is called sulfur dioxide corrosion test chamber
    .

    Temperature error is a comprehensive temperature performance index, it includes: temperature control instrument setting value and the actual temperature deviation in the box, temperature fluctuations, temperature uniformity and other items, the following for you to ask for the salt spray test chamber instrument set value and the actual temperature deviation in the box adjustment method: in the salt spray test chamber there are two temperature control instruments, one control the air temperature in the box, the other controls the water temperature
    in the saturator.

    The setting of these two temperature control instruments has a greater
    impact on the temperature error.

    For the temperature control in the box, the temperature sensor is close to the box wall, which has a distance from the center of the box geometry, so the temperature measured by the temperature sensor, and the temperature of the center point of the box geometry has a difference, the larger the box content, the greater the difference, in order to eliminate the impact of this deviation on the box temperature control, before the formal test, you should first find out this deviation value and correct, so that the temperature measured is close to the center point temperature
    of the box.

     
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.