Small-diameter connectors are devices used for gastrointestinal infusion devices, gastrointestinal syringes, and other devices that are connected to percutaneous intestinal nutrition catheters
.
Medical device manufacturers should provide performance research data
on the size of the connector, fluid leakage, stress cracking, anti-axial load separation, anti-spin-open torque separation, anti-overload, unscrewing separation and non-interconnected characteristics.
The performance tester of small aperture connector is mainly used to determine the physical characteristics of various small aperture connectors, and is one
of the important means to identify small aperture connectors.
Testable includes a variety of 6:100 (ruhr) joint small aperture connections, (locked and non-locked), such as small aperture connections, syringes, infusion sets, blood transfusors, infusion needles, catheters, anesthesia filters, etc
.
YY0916-C small aperture connector performance tester test standard: YY0916-C small aperture connector performance tester is a precision instrument specializing in YY/T 0916.
20-2019 medical liquid and gas small aperture connector Part 20: General test method National standard intelligent precision instrument
.
It can test the fluid leakage, stress cracking, anti-axial load separation, anti-spin-open torque separation, anti-overload, unscrewing separation and non-interconnection characteristics of small aperture connectors, and display the axial force, torque, retention time, internal water pressure, separation force, leakage rate value
of the assembly applied during the test.
Test process: set the parameters, one key can be tested, and the results
are produced in minutes.
With the on-board printing test can display the test data in real time, automatically judge the test results, start the automatic beep alarm if it is unqualified, and print out the test data and results by the printer for analysis
.
Axial force: 1 ~ 100N, the error is not greater than ±0.
5% reading internal water pressure: 200kPa ~ 400kPa, the error is not greater than ±0.
5% reading torque: 0.
01Nm ~ 0.
50Nm, the error is not greater than ±0.
5% Positive pressure leakage rate resolution: 0.
00001Kpa negative pressure leakage rate resolution: 0.
00001Kpa small aperture connector performance tester YYY0916-C small aperture connection performance tester Technical parameters: Programmable controller: PLC+ ARM operating system: Linux operation interface: Chinese Simplified / English touch screen: 7 inch K600 + kernel / dot matrix 65 colors 800* 480 Resolution Sensor: High Precision Force Sensor Printer: On-board needle printer small aperture connector performance tester