Usage method 1, open the box door, put the object to be processed on the shelf in the box, and close the box door
.
2.
Turn on the power supply, insert the three-core plug into the power socket, and put the power switch on the panel in the "on" position, at this time, the digital display of the instrument indicates that the device enters the working state
.
3.
Set the temperature temperature you need by operating the temperature controller on the control panel When the set temperature is greater than the ambient temperature of 5 °C, please place the refrigeration transfer switch in "RT+5 °C"
.
(Detailed operating instructions for the temperature controller can be found on the back page.
)
4, the instrument began to work, the temperature in the box gradually reached the set value, after the required processing time, the processing work was completed
.
Precautions 1, it is strictly forbidden to contain volatile chemical solvents? Explosive and flammable gases are placed in the box, and flammable sprays cannot be used near the incubator? So as not to ignite the electric spark
.
2.
Frequently check whether there is air leakage in the air circuit
.
3, biochemical incubator has a power failure protection function, so the compressor after shutdown to start again to about a minute and a half, so as to better protect the compressor
.
4.
The distance between the condenser and the wall of the biochemical incubator should be greater than 100mm, the side of the cabinet should have a gap of 50mm, and the top of the box should have at least 300mm space to ensure good heat dissipation
.