    Operation procedure: 1, install the injector (clockwise tighten the sampler), the first, second and third stage receiver bottles, open the cooling water circulation button; 2.
    Open the wiper rotor and adjust it to the finger * speed (not more than 400 r/min); 3.
    Open the vacuum pump after the liquid nitrogen reaches the required liquid level (not less than 1/2 of the volume of the condensing column); 4.
    When the pressure no longer drops, adjust the vacuum pump fine-tuning valve (not less than 0.
    5 mbar); 5.
    After the pressure adjustment is in place, rotate the injector valve counterclockwise to adjust the injection speed; 6.
    During the operation of the equipment, the feed liquid can be added at any time through the injector; 7.
    When the liquid volume in the injector remains about 50 ml and is no longer injected, close the vacuum pump, call back the fine adjustment valve, wait for the pressure to show to 1000 mbar, wait for 10 min, wait for the pressure to be discharged, close the heater, remove the injector, pour out the remaining sample solution, and replace it with absolute ethanol; 8.
    Replace the first and second stage receiving bottles, load the empty bottles, unscrew the injector, slowly inject anhydrous ethanol, and borrow waste heat to clean the equipment; 9.
    After cleaning, close the cooling water and rotor, remove the injector and receiver bottle and wash it separately; 10.
    Turn off the power and clean the countertop
    .

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.