    "Serial" paint testing personnel, I would like to say to you (2) - non-volatile content determination, you did the right thing?

    • Last Update: 2019-11-07
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com

    non-volatile content, which is defined as the mass fraction of the residue obtained by the sample by volatile under the prescribed test conditionsThe non-volatile content of the sample is not an absolute amount, the value depends on the measurement of the sample's weighing, the heating temperature and time used and other factorsat present, coatings and related products commonly used in the national standard method is GB / T1725-2007 color paint, varnish and plastic non-volatile content determinationthe test process of common errors are:, the test sample was not fully stirred before sampling
    the sample is not evenly stirred, the sample mentioned as taken is not representative, especially for some easy to layer, precipitation of paint products, such as zinc-rich primer, cloud iron paint, this operation will result in a large deviation in the test results, test results can not be reproducedsaid that when taking samples with glass rods or wooden sticks and other spouts into the test dish
    this operation can not ensure that the sample is evenly distributed, especially for color paint samples containing color fillers, for volatile samples, no weight reduction method is taken to claim the sample
    if the direct weighing method is used to call the volatile sample, the weighing process due to solvent volatilization will lead to inaccurate sample values, resulting in a large deviation in the test resultssaid that after taking the sample did not fully spread the sample in the test flat dish
    the inside of the sample during the baking process containing solventised is not easy to wave, this action will result in high test resultserror diagram
    the above operation is correct:, the test sample is fully stirred evenly immediately called the samplethe sample is called when taking the sample with a spoon (or can ensure that the uniform sample of the apparatus) sampling into the test dish, the volatile sample stirred evenly into the small drop bottle, with the reduction of the amount The method claims to take the sample , said that after taking the sample, the sample is completely or as far as possible spread in the test flat dish
    for high viscosity samples or easy-to-skin samples, using a wire (can be coated with unpainted paperclips) and dish together to weigh the sample When the test accuracy requirements are high, the viscous sample can also be diluted and dispersed with a small amount of thinner the correct picture
    did you do the right thing? relevant wonderful article (click on the title to read) "serial" paint detector, I would like to say to you (1) - from the experimental auxiliary staff W wipe window sill caused by the feeling "guide" paint detector, I would like to say to you (3) - "fine" to operate, reading "degree", how can you do a good job of paint "fineness" test?
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.