    Ruhr joint tester CSI-Z001 Shanghai Chengsi related standards and requirements

    • Last Update: 2022-10-01
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    The first element of the Ruhr joint tester Luer joint is the cone fit, the taper is 6%.


    In order to ensure a strong connection, the Luer joint also derives a second element - thread.


    Threaded, called locking luer joint.


    In practice, most are locking luer fittings
    .

    The Ruhr joint was invented by the Germans, and the inventor was called Luer
    .

    Implementation standard: The international standard for Ruhr fittings is the ISO standard (ISO means the International Committee for Standardization
    ).

    First the ISO 594 series, then the transition to the ISO 80369 series
    .

    The ISO 594 standard was published in its first edition
    in 1967.

    The 2nd edition was divided into two, with Part 1 General Requirements in 1986 and Part 2 Locking Joints
    in 1991.

    ISO 80369 is a larger collection than the Ruhr fitting, of which the part 7 is ISO 80369.
    7
    .

    ISO 80369.
    7 was published in its first edition in 2016 and its 2nd edition
    in 2021.

    In terms of domestic standards, the earliest is the GB 1962 series, and the next step will be to transition to the YY 0916 series
    .

    GB 1962 was published in the first edition in 1980, revised and replaced in 1986 and 1995, and the 4th edition was published in 2001
    .

    Since the 4th edition, the 1962 standard has also been divided into two, and the first part is the general requirements, corresponding to ISO 594-1; Part 2 is to lock the Luer fitting, which corresponds to
    ISO 594-2.

    The first 3 editions are mandatory standards and the 4th edition is recommended
    .

    The YY 0916 standard corresponds to ISO 80369 and consists of 7 parts
    .

    Part 1 is the General Requirements, introduced in 2014; Part 7 corresponds to ISO 80369.
    7 and has not yet been introduced
    .

    The Ruhr joint tester can test the leakage, air leakage, separation force, unscrew torque, easy assembly, anti-slip yarn, stress cracking, positive pressure and negative pressure of the conical joints (locked and non-locking) of medical devices (luer) such as syringes, injection needles, and infusion sets, blood transfusions, infusion needles, catheters, anesthesia filters and other medical device (Luer) cone joints (locked and non-locked
    ).

    Technical parameters: axial force: 20N~40N, error is not more than ±0.
    2N; Internal water pressure: 200kPa~400kPa, error is not more than ±5kPa; Separation force: 20N~40N, error is not more than ±0.
    2N; Hold time: 1~60s, the error is not greater than ±1s arbitrary setting; Torque sensor 0N.
    M-1N.
    M nominal capacity: 1mL ~ 60mL Any choice of power supply: 220V±5% 50Hz This instrument needs to be grounded
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.