    Pses balance LS/LT/LX series fault codes and solutions

    • Last Update: 2022-10-01
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    Electronic balances are precision instruments used only in dry, stable interiors and placed on solid, firm and shock-proof water platforms where excessive fluctuations in airflow and temperature may cause weighing instability
    .

    This article shares the Plyses balance LS/LT/LX series fault codes and solutions
    .

    1, the display is not on ● The balance is not turned on ● The power adapter is disconnected ● There is no power supply (current interruption) ● The power adapter is faulty 2, the display "OL" ● The range is exceeded 3, the "UL" is displayed ● The range is below the range (the weighing pan or weighing pan support is missing) 4, the weight display has been fluctuating ● The airflow at the balance position is large ● The balance support vibrates or the position has been changing ● The weighing pan touches a foreign body ● The floating display selects too short ● The so-called sample absorbs moisture ● The alleged sample is volatilizing or evaporating ● The temperature of the alleged sample is changing greatly5, the weighing result is incorrect ● The balance is not properly peeled to zero ● The balance is not leveled ● Calibration error ● Temperature fluctuates violently6, no display or only a section (————) ● The stability setting is too sensitive (balance function) ● The floating display selection time is not appropriate7, the configuration menu cannot be changed ● The password protection in the configuration menu is activated 8, the display has been flashing during calibration ● The balance placement point is not stable enough (calibration is interrupted with the power key, Then reposition the balance to a better balanced position) ● The calibration weights used are not accurate (external calibration only) and their possible causes are listed
    in the table below.

    If you cannot clear the fault error through this table, please contact us
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.