Note 1. Multi-functional blood plasma machine note 1. Usually set the water temperature, can be long-term start-up, when using the swing switch on, set the frequency, automatic swing, use off the swing switch, save electricity, reduce mechanical wear.
2. Always add water to the water mark before use and do not dry-burn to protect the heating pipe.
as the start-up time continues to increase, the water return of the water tank gradually evaporates, please feel free to check the water level (plus water for water).
ensure that the machine is operating properly.
3. Set temperature: Press SET to set or view the temperature set point.
Press the SET key tube character to start flashing, indicating that the meter into the set state, press the key to set the value to increase, press the key data will change quickly, once again press the SET key meter back to normal working condition temperature set, start heating light.
4. Advance setting: press set key for 3 seconds the meter enters the inner parameter setting state.
The first parameter that appears and flashes is the amount of advance that stops, the advance parameters should be carefully adjusted, in order to reduce the temperature oversleed, the meter controls the heating output will cut off heating ahead of time, when the temperature falls below the advance amount and then start heating, the output (relay) in the set value and the advance quantity range will not work, which can reduce the number of relay actions to extend the life of the relay.
examples: if the setting value is 50.0 degrees C, the advance amount is 0.5, the meter control heating to 49.5.0 when the relay is released, the temperature drops to 50.0 degrees C-0.5 degrees C , 49.5 degrees C when the relay is absorbed again.
the larger the advance, the fewer relay movements, the earlier the amount of the General Assembly to reduce the control accuracy.
the set key for 3 seconds after adjusting the advance parameters, the meter returns to operation.
(note: the advance setting is specially set to prevent the temperature from being too high after the heating stops, and generally does not need to be modified).
5. Correction of error: The display value can be corrected when confirming that the value displayed by the meter is not the correct measurement.
Press the SET key for 3 seconds to enter the instrument deep menu, the first parameter that appears and flashes is E00, that is, the advance quantity, and then press the SET key to flash the parameters, that is, the error correction parameters, with the key to modify this parameter.
the correction range of the error is corrected incorrectly.
is from -9.9 degrees C to 9.9 degrees C, and then press the SET key to exit after the break.
correction value is 0.0 when the meter is factory.
.