    The Prisses XM50 moisture analyzer should be treated with caution and cleaned
    regularly.
    Suzhou Gold Diamond Prisses first-class agent
    .
    To facilitate maintenance work, the device must be completely disconnected from power
    .
    Also make sure that no one else will connect to the power supply during
    maintenance.
    Wash carefully to ensure that no liquid enters the instrument
    .
    If any solution spills into the instrument, cut off the power immediately
    .
    Do not operate the moisture analyzer until it has been checked again by
    a Precisa service engineer.
    The connection connector on the back of the instrument must not come into contact with liquids
    .
    In weighing, regularly remove the weighing pan and tray, remove dirt or dust from the weighing pan, put a soft brush or soft cloth in the weighing balance room, and moisten it with a mild soap solution
    .
    The weighing pan and tray can be cleaned
    with running water.
    When re-inserting the balance, ensure that both are completely dry
    .
    CAUTION: Do not use solvents, acids, alkalis, paint thinners, washout powders, or other "aggressive" or "corrosive chemical cleanings" as these substances may damage the instrument's surface, weighing chamber, and may cause damage
    to the instrument.
    Regular maintenance of your moisture analyzer through Suzhou Gold Diamond Precisa service agent will ensure that the function and reliability of the instrument are not limited for many years, which can extend the life
    of the instrument.
    Replace the power fuse After turning on the instrument, if the display screen is still black, it generally means that the fuse of the instrument is wrong and needs to be replaced
    .
    Before replacing the fuse, unplug the power supply and use a screwdriver to turn counterclockwise to open the fuse retainer on the back of the instrument
    .
    Replace the wrong fuse: – 230 volts: t 3.
    15 A, 230 V, 5x20 mm – 115 volts: t 6.
    3 A, 115 V, 5x20 mm If the instrument still does not work after replacing the fuse, contact the Suzhou Gold Diamond Precisa service agent
    .
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.