Operation specifications of single person purification workbench: 1.
Check the operating environment of single person clean workbench such as ambient temperature, air quality, power supply, etc.
, and register the use information; 2.
Place the equipment required for the experiment on the clean workbench, put sterilized materials such as culture medium on the left side, and put the tools on the right; 3.
Pull down the glass baffle and turn on the UV lamp irradiation for 30 minutes for surface sterilization; 4.
Turn off the UV lamp after 30 minutes, turn on the fan to adjust to the appropriate air volume, and turn on the lighting; 5.
Open the glass baffle to about 20cm high and two hands can be operated smoothly in the operating table, wipe the hands and table with 75% alcohol; 6.
Light the alcohol lamp, and strictly follow the aseptic operation specifications during the operation process; 7.
After the operation is completed, clean the experimental equipment out of the aseptic operation table, wipe the table with 75% alcohol, and turn off the lighting, fan and baffle; 8.
Turn off the UV lamp after turning on the UV lamp
for 15 minutes.
Second, the management specifications of the single purification workbench: 1.
Every time the clean workbench is used, the experimenter should first turn on the ultraviolet lamp on the clean workbench, and use it after 20 minutes of ultraviolet irradiation; 2.
Turn on the working power of the clean workbench, turn off the UV lamp, and wipe and disinfect the work surface with 75% alcohol or 0.
5% peracetic acid spraying; 3.
During the whole experiment, the experimental personnel should operate in accordance with the aseptic operating procedures; 4.
After the experiment, wipe the work surface with disinfectant, turn off the working power, and re-turn on the ultraviolet lamp for irradiation; 5.
In case of a failure of the unit, the animal laboratory should be notified immediately, and it will continue to be used after the personnel have passed the maintenance; 6.
Experimenters should pay attention to keeping the room tidy; 7.
The filter material of the single clean table is replaced once every 2 years, and a replacement record
is made.