echemi logo
Product
  • Product
  • Supplier
  • Inquiry
    Home > Active Ingredient News > Drugs Articles > Inhalation preparation drug tester Shanghai Chengwei

    Inhalation preparation drug tester Shanghai Chengwei

    • Last Update: 2022-10-02
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    Inhalation preparation drug tester one, the implementation standard in line with YY/71743-2021 for the instrument of the standard two, technical parameters 1.
    The flow rate from 15L/min to 100L/min adjustable; 2.
    Contains 7 layers of impactor and a micro-hole collector; 3.
    With 8 movable collection cups; 5.
    The impactor contains a pre-separator, which can be used to detect inhaled aerosols, inhaled sprays, inhaled powder aerosols and inhaled liquid preparations
    .

    The pre-separator can effectively remove large particles of 10-15 μm generated by the dry powder inhaler; 6.
    The collection cup is the impact surface, placed on the tray, directly below the nozzle; 7.
    The dose uniformity test
    of inhaled powder aerosol, inhaled aerosol and inhaled spray can be completed.

    8.
    The artificial throat is made of stainless steel, and the size meets the requirements
    of the pharmacopoeia.

    9.
    The pre-separator is made of stainless steel, with a pore diameter of 12.
    80±0.
    05mm, which meets the requirements
    of the pharmacopoeia.

    10.
    First-stage nozzle, pore diameter 14.
    30±0.
    05mm, in line with the requirements
    of the pharmacopoeia.

    11.
    Secondary nozzle, pore diameter 4.
    88±0.
    04mm, in line with pharmacopoeia requirements
    .

    12.
    Three-stage nozzle, pore diameter 2.
    185±0.
    02mm, in line with the requirements
    of the pharmacopoeia.

    13.
    Four-stage nozzle, pore diameter 1.
    207±0.
    01mm, in line with pharmacopoeia requirements
    .

    14.
    Five-stage nozzle, pore diameter 0.
    608±0.
    01mm, in line with pharmacopoeia requirements
    .

    15.
    Six-stage nozzle, pore diameter 0.
    323±0.
    01mm, in line with the requirements
    of the pharmacopoeia.

    16.
    Seven-stage nozzle, pore diameter 0.
    206±0.
    01mm, in line with pharmacopoeia requirements
    .

    17.
    MOC nozzle, pore size of about 0.
    07mm, in line with pharmacopoeia requirements
    .

    18.
    The small collection cup is made of stainless steel, the depth value is 14.
    625±0.
    1mm, and the surface roughness (Ra) is 0.
    5~2μm in line with the requirements
    of the pharmacopoeia.

    19.
    The large collection cup is made of stainless steel, the depth value is 14.
    625±0.
    1mm, and the surface roughness (Ra) is 0.
    5~2μm in line with the requirements
    of the pharmacopoeia.

    20.
    Intelligent pump, the flow rate is adjustable from 0 to 90L/min, the pumping time is adjustable from 0.
    1S to 9999S, and the noise ≤ 75dB
    .

    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.

    Contact Us

    The source of this page with content of products and services is from Internet, which doesn't represent ECHEMI's opinion. If you have any queries, please write to service@echemi.com. It will be replied within 5 days.

    Moreover, if you find any instances of plagiarism from the page, please send email to service@echemi.com with relevant evidence.