The light incubator has ultra-temperature and sensor abnormal protection function to protect the instrument and sample; Optional full-spectrum plant growth lamp, which is conducive to plant growth and improves disease resistance
.
With power loss memory, power down time automatic compensation function; Constant temperature control system, fast response, high
temperature control accuracy.
Light incubator microcomputer automatic control, touch switch, easy to operate; Programmable multi-segment control mode, day and night can be set separately temperature, humidity, illuminance and time
.
Wind duct ventilation, soft air speed in the studio, uniform temperature; Hollow reflection tempered film glass, good thermal insulation performance, beautiful and generous
.
Fully enclosed opaque lampshade, optional studio power supply, installation, etc
.
Maintenance and incubation1, the incubator housing should be reliably grounded
.
2.
The incubator should be placed in a cool, dry, well-ventilated place
, away from heat sources and sunlight.
Position smoothly to prevent vibration from noise
.
3.
In order to ensure the heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm
.
There should be a 50mm gap on the side of the box, and at least 300mm space
on the top of the box.
4.
When the incubator is handled, repaired and maintained, collisions and shaking should be avoided; The large inclination is less than 45 degrees
.
5, the instrument suddenly does not work, please check whether the fuse pipe (behind the box) burned out, check the
power supply.
6.
When the incubator is refrigerated, it is not advisable to make the difference between the temperature in the box and the ambient temperature greater than 25 degrees
.