Ship fuel oil GB17411-2015 in the kinematic viscosity detection method Ship fuel oil GB17411 in the kinematic viscosity detection method in accordance with GB/T265 to determine, in line with the GB/T265 standard instrument is generally automatic motion viscosity meter SH112C, through this to determine the 40 degrees of kinematic viscosity
of ship fuel oil.
The fully automated kinematic viscosity analyzer SH112C is suitable for measuring the kinematic viscosity
of samples with high wax content or samples containing insoluble components at room temperature.
Automatic injection, constant temperature, viscosity testing, cleaning, drying viscometer, no need for random operation of personnel, the operator can leave the site after lofting, and the instrument can automatically complete all tasks
.
According to the standard requirements, the kinematic viscosity of marine fuel oil is also tested
at a constant temperature of 40 degrees using a Pinnacle viscometer.
Pinnacle viscometers should be cleaned with solvent oil or petroleum ether, and cleaned
sequentially with ethanol and water.
If there is a residue attached to the glass wall, it can be cleaned with chromic acid solution or other strong oxidizing acid lotion without chromium, and then cleaned with solvent oil or petroleum ether, ethanol and water in turn, and dried with dry hot air or upside down in the oven at a temperature of 100 ° C ~ 110 ° C for more than
1h.
Pinnacle viscometer should comply with the provisions of BMN-1 and BMN-2 in SH/T 0173 "Technical Conditions for Glass Capillary Viscometers"
.
The necessary packaging and measures should be taken to prevent the viscometer from breaking
in transportation and storage.