-
Categories
-
Pharmaceutical Intermediates
-
Active Pharmaceutical Ingredients
-
Food Additives
- Industrial Coatings
- Agrochemicals
- Dyes and Pigments
- Surfactant
- Flavors and Fragrances
- Chemical Reagents
- Catalyst and Auxiliary
- Natural Products
- Inorganic Chemistry
-
Organic Chemistry
-
Biochemical Engineering
- Analytical Chemistry
-
Cosmetic Ingredient
- Water Treatment Chemical
-
Pharmaceutical Intermediates
Promotion
ECHEMI Mall
Wholesale
Weekly Price
Exhibition
News
-
Trade Service
The catheter negative pressure tester adopts 7-inch color touch screen, Chinese and English menu display
.
And by the touch screen to select the nominal capacity of the product, display mode: can arbitrarily set the pressure output value, can print out the report, display time adjustable, suitable for the development of products used in the equipment
.
The verification error is not greater than 1%
of the ± of the reading.
Implementation standards: in line with YY0285.
1-2017 Appendix C standards in line with YY0267-2016, GB/T15812.
1-2005, YY02851-2017 in the negative pressure resistance requirements
.
Technical parameters: 1, test flow: 30L/min, error is not more than ±5%2, pressure range: -100kpa to 500KPa3, resolution: 0.
01 KPa, error is not more than ±1% 4, measurement packing time: 0-99999 seconds, can be arbitrarily set 5, test duration: 1 second ~ 99.
9 minutes, error is not more than ±1S
.
6.
Suitable for infusion sets, blood transfusion devices, infusion needles, anesthesia bag filters, pipelines, catheters, quick couplings, etc
.
7, flow leakage detection sensor: measuring range 0-50mL/min, accuracy ± 5% 8, reagent: steamed stuffing water or deionized water
.
Configuration list: 2 sets of Luer fittings, one pair of hemostatic forceps