    Common fault inspection methods of electric vacuum drying oven: 1.
    The vacuum cannot be pumped up 1.
    The sealing strip is aging and ineffective; Replace the sealing strip, and try not to contact the sealing strip with organic solvents
    .

    2.
    The vacuum pump fails; replace the vacuum pump
    .

    3.
    The door is not closed tightly; close it again and push the glass lightly when starting to vacuum
    .

    4.
    The vacuum degree of the vacuum gauge shows the illusion; replace the vacuum gauge
    .

    Second, the temperature fluctuates greatly 1.
    The external power supply voltage changes greatly; Take voltage stabilization measures to make it meet the power supply requirements of the equipment
    .

    2.
    The temperature control instrument drifts; replace the temperature control instrument
    .

    3.
    No temperature rise 1.
    The heater is open circuit; after eliminating the possibility of wiring open circuit, replace the heater if the fault remains
    .

    2.
    The sensor is open circuit; after eliminating the possibility of wiring open circuit, replace the sensor if the fault remains
    .

    3.
    The relay is open circuit; after eliminating the possibility of wiring open circuit, replace the relay if the fault remains
    .

    4.
    The temperature control instrument is not set well; reset the temperature control instrument according to the requirements of the manual
    .

       Fourth, the temperature control fails and the temperature rises 1.
    The sensor is short-circuited; after eliminating the possibility of the wiring short-circuit, replace the sensor if the fault persists
    .

       2.
    The relay is short-circuited; replace the relay
    .

       3.
    The temperature control instrument is invalid; replace the temperature control instrument
    .

     
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.