The debugging method of automatic liquid filling machine is automatic quantitative liquid dispensing machine designed by electric, crank, piston type, peristaltic pump and other structures, which is suitable for the quantitative filling
of hospital preparation room, ampoule, eye drops, various oral liquids, syrups and various water agents.
The debugging method of the automatic liquid filling machine has the following points: 1, first loosen the nut and then adjust the screw so that the position of the lower fixing rod changes
.
Thus, the push and pull length of the syringe has also been changed, so as to achieve the purpose
of free adjustment of the filling amount.
The screw is adjusted clockwise to increments and vice versa, with the pointer indicated
.
With a 5ml syringe, look at the 5ml side of the label, use the 10ml needle tube to see the label of 10ml while directly looking at the ruler, the pointer quantification, but because the thickness of each needle tube is different, the actual amount of filling has an error, subject to the needle tube scale, you can also use the measuring cup to measure the most accurate
when starting to dispense.
2.
Immediately after the joint, the nut is tightened so that the lower fixing rod and crank are
firmly fixed.
3.
Loosen the nut and determine the correct position
of the upper fixing rod according to the push and pull length of different syringes.
After the lower fixing rod is adjusted, turn the crank to turn the syringe to the upper dead point, and then lift the syringe manhole tube up about 2mm to avoid ejecting the needle tube, and then screw the nut tightly before it can be turned
on.
4, the machine assembly is error-free after opening the switch, when the machine is working, the crank drives the syringe up and down to pull the pump, adjust the speed knob, choose the appropriate sub-assembly speed, start normal work, adjust the governor knob, clockwise sub-assembly speed is fast, vice versa speed is slow
.