    Home > Active Ingredient News > Drugs Articles > Combined oscillation incubator maintenance

    Combined oscillation incubator maintenance

    • Last Update: 2022-09-15
    • Source: Internet
    • Author: User
    Search more information of high quality chemicals, good prices and reliable suppliers, visit www.echemi.com
    Before the initial use, the combined oscillation incubator should first use the thermostat power plug with a multimeter for a safety inspection, with one of the measuring resistors, the amount of plug phase, middle, ground, whether there is a short circuit or bad insulation between each other; Add distilled water (water level about 30-43 mm from the cover plate) according to the regulations to connect the power plug to the power supply, open the power switch and electric pump switch on the control box, and make the water in the tank circulate convection; Adjust the thermostatic water bath to the set temperature
    .

    Combined oscillation incubator maintenance 1, the compressor should avoid continuous starting 2, the bearing of the transmission part has been filled with an appropriate amount of grease before leaving the factory, the instrument should be filled with grease every six months during continuous operation, and the filling amount accounts for about 1/3
    of the bearing space.

    3.
    During continuous operation of the oscillating incubator, it should be checked regularly every three months; Check the fuse, control components and fastening screws, and whether water droplets, dirt, etc.
    fall onto the
    motor and control elements.

    4.
    The oscillation incubator is often used, and natural wear and tear is a normal phenomenon
    .

    After one year of use, if the instrument is found to have abnormal noise in the motor and compressor; Transmission part of the bearing wear, belt loose or cracked, electronic control components failure and other failures, the enterprise will continue to provide services, to assist in the treatment
    .

    5.
    The oscillation incubator should be placed on a flat ground, and the environment should be clean and tidy, dry and ventilated
    .

    6.
    Before use, each control switch should be in a non-working state, and the speed control knob should be placed in a small position
    .

    7, loading culture test bottles, in order to make the combined oscillation incubator in the oscillation balance performance is good, to avoid larger vibration, bottle loading should be filled with all the test bottles, each bottle of culture liquid should be roughly equal
    .

    If there is not enough culture solution, the test bottle can be placed symmetrically
    .

     
    This article is an English version of an article which is originally in the Chinese language on echemi.com and is provided for information purposes only. This website makes no representation or warranty of any kind, either expressed or implied, as to the accuracy, completeness ownership or reliability of the article or any translations thereof. If you have any concerns or complaints relating to the article, please send an email, providing a detailed description of the concern or complaint, to service@echemi.com. A staff member will contact you within 5 working days. Once verified, infringing content will be removed immediately.