1.
Environmental requirements 1.
Ambient temperature: 15°C~35°C; 2.
Relative humidity: not more than 85%RH (at 25 °C); 3.
Power supply voltage: AC198~222V, 50/60Hz; 4.
There is no strong light around, no corrosive gas, good ventilation, no strong vibration source and strong electromagnetic field
.
Second, the incubator panel diagram (1) operation and use method (2) Intelligent controller(3) Heating indicator (4) Cycle indicator (5) Lighting switch (6) Power switch (7) Transfer switch (8) Refrigeration indicator 3.
Operation step 1, put the items that need to be used for culture experiments into the incubator working room, and leave space around the upper and lower sides to keep the air flowing in the working room and close the door
.
2.
Turn on the power switch, at this time the cycle indicator light is on, the motor is running, the PV screen of the intelligent controller should display the measured temperature in the working room, and the SV screen should display the set temperature
to be used.
The incubator is now in working condition
.
3.
Set the desired temperature: press the SET button once, at this time the PV screen displays the "SP" character, the SV screen * bit is highlighted, and other bits are flashing, use the cycle to select the specific bit of the highlight displayed on the SV screen, and use or change the temperature value displayed on the original SV screen until the value is needed
.
After setting, press the SET key again, PV displays the "TI" character, if the timing function is not used, press the SET button again to make the PV screen display measurement temperature, SV screen display set temperature can be
.
Note: If you do not use the timing function, you must display the "TI" character on the PV screen, and the SV screen is displayed as "0" 4.
Set the timing time: see "How to use the timing function" 5.
Select the transfer switch: a.
When the setting temperature is 5 °C higher than the ambient temperature, please select the "RT+5°C" file; b.
To set the temperature below the ambient temperature +5°C or close to the ambient temperature, select the "Auto" setting
.
6.
When the machine is turned on, or used for a period of time, or when the season (ambient temperature) changes, the measured temperature and the actual temperature error in the working room must be reviewed, and the specific method is shown in the "Instructions for the Use of Each Function Parameter"
.
7.
After each use, be sure to dry the water in the studio and turn off the main power to ensure that
.