Anesthesia puncture catheter bending resistance tester one, instrument features: This equipment is an efficient testing equipment
for national inspection of the product performance of manufacturers.
It is composed of programmable controller, touch screen, force sensor, transmission device, airborne printer, etc.
, provides Chinese and English menu display, and sets the automatic operation test mode
of various parameters of man-machine dialogue.
In addition, production is added, and the production batch machine is input to make the test results more standardized and standardized
.
In terms of information security, three-level passwords are adopted to make information safe and reliable
.
The test control range includes: pressure, printing settings, testing, uplink, downlink, time, calibration and other function settings
.
In addition, it is equipped with an on-board printer to display test results and load and disturbance data
in real time.
Automatic judgment of test results, automatic alarm for failure to conform, and print out test data and results by the printer for analysis
.
It is an effective testing method
for the state to inspect the product capacity of manufacturers.
Second, applicable standards: YY/T0321.
1-2022 and other related standards 3, technical parameters 1, pressure sensor range: 0-500kpa; 2, PLC control system; 3, medical testing software 1 set; 4, built-in 2-liter stainless steel air tank; 5, embedded micro printer with printing paper; 6, measuring cup 50ML---1 piece; 7, operation interface: Chinese and English switching; 8, color 7-inch touch screen, atmosphere, beautiful; 9, Luer connector: lock Luer connector 2 sets; IV, Test step 1, turn on the power of
the instrument.
2.
The catheter to be tested is installed on
the outer cone joint of the water outlet of the instrument.
3.
Adjust the water pressure to the standard specified value
.
4.
Select the test time
.
5.
Start the test switch and let the water be injected into the measuring cylinder from the outlet end of the catheter under test.
Automatically closes
when the set is reached.
6.
Read the water volume in the
measuring cylinder.
7.
Bend the conduit according to the method of Figure E.
1 and put it into a slot with 2 times the diameter width and repeat the test once
.