The high and low temperature alternating humidity and heat test chamber is mainly used to test the weather resistance adaptability of consumer products, simulate the climatic and geographical environment such as high temperature, low temperature, high temperature, high and low temperature tests, and low temperature and low temperature drying, so as to identify the efficiency of products and the development of new products at that time.
defect
.
It is an indispensable testing equipment to improve the core competitiveness of goods.
It is used in hardware, plastics, electronic components, toys, lighting equipment, LED, motherboard interfaces.
Rationalized experimental instrument for quality inspection
.
In the normal application process, some consumers are not very skilled in the application, resulting in the situation that the high and low temperature alternating humidity and heat test box sometimes causes the power failure after starting up.
Questions from all walks of life
.
1.
If this happens, check one by one, first set the machine and equipment to only high temperature (40 degrees Celsius or 60 degrees Celsius), set the relative humidity of the air to 0, click to run, and see if the machine equipment will trip, it is the heating tube Short-circuit fault problem, resulting in electric shock accident, high and low temperature alternating humidity and heat test box will cause all aspects of maintenance will cause electric shock
.
In this case, the short-circuit fault of the heating tube can be replaced by replacing the lower heating tube, and the operation process is very simple
.
2.
If the operation does not trip at high temperature, set the relative humidity of the air, and then click Run again to see if the high and low temperature alternating damp heat test box is tripped
.
If it is tripped, it is a short circuit fault of the humidifier tube, and the electricity is disconnected
.
If this happens, just unscrew the baffle screw on the side of my driver's device, take off the baffle, you will see a humidifier bucket, remove the two wires on it, and remove the rear cover.
Remove the 6 socket head cap bolts and replace the humidifier tube.
At the time of installation, be sure to connect the temperature sensor monitoring camera to the humidifier tube to prevent empty burning and over-temperature maintenance
.
defect
.
It is an indispensable testing equipment to improve the core competitiveness of goods.
It is used in hardware, plastics, electronic components, toys, lighting equipment, LED, motherboard interfaces.
Rationalized experimental instrument for quality inspection
.
In the normal application process, some consumers are not very skilled in the application, resulting in the situation that the high and low temperature alternating humidity and heat test box sometimes causes the power failure after starting up.
Questions from all walks of life
.
1.
If this happens, check one by one, first set the machine and equipment to only high temperature (40 degrees Celsius or 60 degrees Celsius), set the relative humidity of the air to 0, click to run, and see if the machine equipment will trip, it is the heating tube Short-circuit fault problem, resulting in electric shock accident, high and low temperature alternating humidity and heat test box will cause all aspects of maintenance will cause electric shock
.
In this case, the short-circuit fault of the heating tube can be replaced by replacing the lower heating tube, and the operation process is very simple
.
2.
If the operation does not trip at high temperature, set the relative humidity of the air, and then click Run again to see if the high and low temperature alternating damp heat test box is tripped
.
If it is tripped, it is a short circuit fault of the humidifier tube, and the electricity is disconnected
.
If this happens, just unscrew the baffle screw on the side of my driver's device, take off the baffle, you will see a humidifier bucket, remove the two wires on it, and remove the rear cover.
Remove the 6 socket head cap bolts and replace the humidifier tube.
At the time of installation, be sure to connect the temperature sensor monitoring camera to the humidifier tube to prevent empty burning and over-temperature maintenance
.