The constant temperature shaking incubator is a special constant temperature cultivation device for mold and microorganism cultivation and breeding experiments
.
It is also a two-way constant temperature product with both cooling and heating, and with oscillation.
Because of its complete functions, it is also widely used
.
Constant temperature shaking incubators are used for dynamic culture in plants, organisms and microorganisms
.
When there is a sound during the use of the constant temperature shaking incubator, you can deal with it as follows: 1.
You should first rule out the placement of the whole machine in a flat position before checking other parts of the problem
.
2.
Check the constant temperature part and close the oscillation part.
When the constant temperature oscillating incubator is refrigerating, the compressor will make a sound, and the refrigerant in the compressor will also make a sound when it flows in the refrigeration pipe, but it is normal if the sound is not harsh, such as There is a harsh sound, a crashing sound, and this should be further checked
.
3.
The abnormal sound of the constant temperature oscillating incubator generally comes from the oscillating part.
Because the oscillating part of the incubator is inevitably worn for a long time, it will cause abnormal sound.
When checking the oscillating part of the incubator, the constant temperature part should be closed, and the oscillation part should be closed.
Adjust the speed to a slow speed and carefully observe where the sound comes from.
If the bearing loses oil, remove the rocking plate, add less lubricating oil to the bearing with oil loss, and then restore the oscillation part
.
.
It is also a two-way constant temperature product with both cooling and heating, and with oscillation.
Because of its complete functions, it is also widely used
.
Constant temperature shaking incubators are used for dynamic culture in plants, organisms and microorganisms
.
When there is a sound during the use of the constant temperature shaking incubator, you can deal with it as follows: 1.
You should first rule out the placement of the whole machine in a flat position before checking other parts of the problem
.
2.
Check the constant temperature part and close the oscillation part.
When the constant temperature oscillating incubator is refrigerating, the compressor will make a sound, and the refrigerant in the compressor will also make a sound when it flows in the refrigeration pipe, but it is normal if the sound is not harsh, such as There is a harsh sound, a crashing sound, and this should be further checked
.
3.
The abnormal sound of the constant temperature oscillating incubator generally comes from the oscillating part.
Because the oscillating part of the incubator is inevitably worn for a long time, it will cause abnormal sound.
When checking the oscillating part of the incubator, the constant temperature part should be closed, and the oscillation part should be closed.
Adjust the speed to a slow speed and carefully observe where the sound comes from.
If the bearing loses oil, remove the rocking plate, add less lubricating oil to the bearing with oil loss, and then restore the oscillation part
.